Calibration
4-1
Calibration
Introduction
The LI-7500’s measurement accuracy depends upon its calibration. There are two major
components to the calibration: 1) determining the values of calibration coefficients, and 2)
setting zero and span. During a factory calibration, both of these steps are performed. The
values of the coefficients that are determined should be valid for several years. The zero and
span adjustments are used to bring the LI-7500's actual response into line with its previously
determined factory response, at least at two points. It is the user's responsibility to do this step
at regular intervals (weekly or monthly).