Section 6
Troubleshooting
6-3
Bad Temperature or Pressure Readings
The source of the values of temperature and pressure displayed on the Calibration page are
determined by the setting on the Inputs page. If the source is "Measured", the readings come
from the LI-7500's on-board sensors (or the thermistor on the calibration tube, if that is
connected). If the source is "Aux Input", then the temperature or pressure (but not both) are
coming from an external sensor that you have wired and connected. This signal is modified
according to the Auxiliary Coefficients that appear on that same page. If the source is "User-
Entered", then whatever you enter in the text boxes is used for the value.
Temperature is Measured, and reads -65 (or so)
The thermistor is broken, or disconnected. Check the 6 pin connector on right of the board to
make sure it is plugged in, and to make sure the thermistor wires are both attached to the
connector at pins 4 and 5.
Pressure is Measured, and doesn't match my barometer
The pressure sensor in the LI-7500 is good to about 1%. For operating purposes, it doesn't
need to be very good (See "Note on Temperature and Pressure" on page 2-12). When setting
the span of the LI-7500, however, it is important. If you know the pressure better than the
LI-7500 does, then use the "User-Entered" option and enter it.
Содержание LI-7500
Страница 1: ...Analyzer Environmental Division Instruction Manual 7500...
Страница 2: ...Open Path CO2 H2O Analyzer Instruction Manual...
Страница 41: ...Section 3 3 12 Setup and Operation...
Страница 43: ...Section 3 3 14 Setup and Operation...
Страница 46: ...Section 3 Setup and Operation 3 17...
Страница 50: ...Section 3 Setup and Operation 3 21 Inputs Page Temperature Pressure and Auxiliary Inputs...
Страница 56: ...Section 3 Setup and Operation 3 27...
Страница 58: ...Section 3 Setup and Operation 3 29...
Страница 67: ...Section 3 3 38 Setup and Operation...
Страница 70: ...Section 3 Setup and Operation 3 41...
Страница 107: ...Section 6 6 6 Troubleshooting...