Section 3
3-26
Setup and Operation
Auxiliary Input Example:
Suppose you have a custom temperature sensor that outputs a linear 0 to +5VDC signal. This
signal represents a -40 °C to +60 °C temperature range. The LI7500 software allows you to
enter the proportional multiplier (slope) and offset for this sensor. (Select the Temperature
Aux Input radio button of the Inputs Tab to activate this selection).
The LI7500 software assumes you are entering a slope in units of degrees (or kPa) per VOLT.
In the example using an external temperature sensor listed above we would calculate the slope
by taking the output range of the sensor divided by the proportional output voltage:
− °
° = °
=
°
40 C to + 60 C
0 to + 5V
100 C Range
5V
20 C
Volt
This value (20) is then entered into the auxiliary input coefficient section. Since the output
range goes below zero, a value of -40 is entered into the offset section.
Содержание LI-7500
Страница 1: ...Analyzer Environmental Division Instruction Manual 7500...
Страница 2: ...Open Path CO2 H2O Analyzer Instruction Manual...
Страница 41: ...Section 3 3 12 Setup and Operation...
Страница 43: ...Section 3 3 14 Setup and Operation...
Страница 46: ...Section 3 Setup and Operation 3 17...
Страница 50: ...Section 3 Setup and Operation 3 21 Inputs Page Temperature Pressure and Auxiliary Inputs...
Страница 56: ...Section 3 Setup and Operation 3 27...
Страница 58: ...Section 3 Setup and Operation 3 29...
Страница 67: ...Section 3 3 38 Setup and Operation...
Страница 70: ...Section 3 Setup and Operation 3 41...
Страница 107: ...Section 6 6 6 Troubleshooting...