Section 6
6-8
Troubleshooting
Absorptance values - CO
2
: computed according to Equation 2-16.
H
2
O: computed according to Equation 2-12.
Cooler Voltage - Displays voltage of detector cooler. If this value gets below 0.05, it may be
too cold; the detector may not be temperature controlling. This will typically be at -30 °C or
below.
AGC - Note this value when you first operate the LI-7500. Typical values are 50-60%. As
dirt accumulates on the sensor head windows the value of AGC will increase, so you can use
this value to indicate when it is necessary to clean the windows.