Appendix A
A-1
Specifications*
Type: Absolute, open-path, non-dispersive infrared gas analyzer.
Detector: Thermo-electrically cooled lead selenide.
Bandwidth: 5, 10, or 20 Hz, software selectable.
Path Length: 12.5 cm (4.72").
Operating Temperature Range: -25 to 50°C (-40 to 50 °C verification test available on request).
User Interface: Windows
®
based software supports all setup, configuration and calibration functions through RS-232 serial port.
Outputs: RS-232 (20 Hz Maximum)
SDM (>40Hz)
2 user scalable 16 bit DACs updated at 300 Hz.
Auxiliary Inputs: 2 channels for temperature and pressure sensors (during calibration).
Auxiliary Input with pressure sensor: 0-4.096V (±5V common mode rejection).
Power Requirements: 10.5 to 30 volts DC.
Power Consumption: 30 W during warm-up, 10 W in steady state.
Dimensions:
Head: Dia 6.5 cm, Length 30 cm. Designed for minimal wind flow disturbance.
Содержание LI-7500
Страница 1: ...Analyzer Environmental Division Instruction Manual 7500...
Страница 2: ...Open Path CO2 H2O Analyzer Instruction Manual...
Страница 41: ...Section 3 3 12 Setup and Operation...
Страница 43: ...Section 3 3 14 Setup and Operation...
Страница 46: ...Section 3 Setup and Operation 3 17...
Страница 50: ...Section 3 Setup and Operation 3 21 Inputs Page Temperature Pressure and Auxiliary Inputs...
Страница 56: ...Section 3 Setup and Operation 3 27...
Страница 58: ...Section 3 Setup and Operation 3 29...
Страница 67: ...Section 3 3 38 Setup and Operation...
Страница 70: ...Section 3 Setup and Operation 3 41...
Страница 107: ...Section 6 6 6 Troubleshooting...