Appendix D
D-8
Appendix D
Example:
Transmitted:
(Outputs(RS232(DiagVal TRUE)))lf
Received:
(Data... (Diag 249...))
NOTE: The (DiagRec and the (DiagVal are two separate outputs and are independent of one
another.
(Labels {bool})
This command controls whether or not data labels are transmitted with the data stream.
Default is TRUE, and this means that data are transmitted in the normal (Data) format record.
When (Labels FALSE), however, data output are values only and are tab-delimited. To
maximize available bandwidth, when (Labels FALSE), a data record might appear as
15.1234<tab>354.123<tab>101.3<tab>249<eol>
The following are several examples of data formatting. The command string sent to the LI-
7500 is terminated with a line feed character.
EXAMPLE (Labels TRUE) Data format
Transmitted to LI-7500
(Outputs(RS232(Freq 1)(EOL "0D0A")(Ndx TRUE)(DiagRec FALSE)(DiagVal
TRUE)(Labels TRUE)))lf
Received from LI-7500