Troubleshooting
6-1
Troubleshooting
Power On Problems
The instrument requires between 10 and 30 VDC, at 2 Amps, to operate (a LI-COR LI-6020
battery charger, for example, will not by itself power the LI-7500). When adequate power is
first applied to the instrument, the LED in the box near the fuse (Fig 5-1) will light up for
about 5 seconds, then (if operating correctly) go off and remain off. If any of the diagnostics
listed below become true during the operation of the instrument, then the LED will come back
on.
NOTE: On instrument serial numbers 75B-0370 and above, a red LED is used as the status
indicator, which should turn off after 5-10 seconds. A second green LED should stay lit when
the LI-7500 is powered on.
LED does not light at power on