![LI-COR LI-7500 Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/li-cor/li-7500/li-7500_instruction-manual_1906655023.webp)
Section 2
2-6
Theory of Operation
Equation Summary
H
2
O
In the atmosphere, the absorption of radiation by water vapor is not significantly influenced
by any other gas, so the effective pressure for water vapor P
ew
is simply the total pressure P.
P
ew
= P
2-11
H
2
O absorptance
α
w
is (from Equations 2-8 and 2-10)
α
w
w
wo
cw
c
co
wo
w
d
A
A
X
A
A
Z
Z V
= −
+
−
+
(
)
1
1
2-12
where X
cw
is the cross sensitivity factor for CO
2
on water vapor ("H
2
O XS" on the calibration
sheet), Z
w
is the zero drift coefficient ("H
2
O Z" on the calibration sheet).
H
2
O number density
ρ
w
is given by
ρ
α
w
w
w
w
Pf
S
P
=
2-13
The coefficients for the 3rd order polynomial f
w
( ) are given on the calibration sheet.
Содержание LI-7500
Страница 1: ...Analyzer Environmental Division Instruction Manual 7500...
Страница 2: ...Open Path CO2 H2O Analyzer Instruction Manual...
Страница 41: ...Section 3 3 12 Setup and Operation...
Страница 43: ...Section 3 3 14 Setup and Operation...
Страница 46: ...Section 3 Setup and Operation 3 17...
Страница 50: ...Section 3 Setup and Operation 3 21 Inputs Page Temperature Pressure and Auxiliary Inputs...
Страница 56: ...Section 3 Setup and Operation 3 27...
Страница 58: ...Section 3 Setup and Operation 3 29...
Страница 67: ...Section 3 3 38 Setup and Operation...
Страница 70: ...Section 3 Setup and Operation 3 41...
Страница 107: ...Section 6 6 6 Troubleshooting...