Setup and Operation
3-1
Setup and Operation
Installing the LI-7500
Mounting the Control Box
Brackets are included that can be used to mount the control box using bolts or U-clamps to
secure the box to a tripod or other post. There are holes in the four corners of the box, as well,
that can be used to attach the box directly to a flat surface, if desired.
The mounting brackets are designed to accomodate a square or circular mounting post of up to
1 1/4" in width (diameter). Use a 3/16" hex key to attach the mounting brackets to the control
box. The U-bolt can then be inserted through the holes in the mounting bracket, and tightened
around the post, as shown in Figure 3-1.