If necessary, the device carries out the degassing automatically. If the device
detects gas in the hydraulic system, the heating and cooling output is first
reduced or temporarily switched off completely. If the pump pressure drops
sharply (a clear sign of degassing), the pump speed is limited and the pump
may be switched off briefly. When the automatic degassing is finished, the
device resumes its operation automatically.
5.7 Setting the flow rate of the heat transfer liquid
CAUTION!
Bursting of the external application due to overpressure
Scalding, frostbite, impacts
For applications with a maximum permissible working pres-
sure that is less than the maximum pressure of the pump,
use a pressure relief device for protection. This pressure
relief device must be installed in the outflow to the applica-
tion.
Set the maximum pump pressure as per your application.
This section is relevant for:
n
Integral T devices
Before switching on, turn the bypass control knob counterclockwise
to the fully open position.
Personnel:
n
Operating personnel
Protective equipment:
n
Safety glasses
n
Protective gloves
n
Protective work clothing
In the Integral T devices, the flow rate is set via the bypass on the right-hand
side of the device. The bypass enables operation with a pressure-sensitive
application.
1.
To increase the flow rate and thus the pressure in the consumer, turn
the bypass adjusting wheel clockwise until the desired (< permissible)
pressure for the external consumer is reached.
Observe the changing display of
[pump pressure]
in the
display.
To display the pump pressure, select the menu items
Pump
Pump
pressure
.
The current pressure of the pump in the outflow is displayed. No
settings can be made here.
Automatic degassing
Setting flow rate for Integral T
V6
Integral Process Thermostats and High-Temperature Thermostats
77 / 198
Содержание IN 1030 T
Страница 89: ...6 3 Menu structure Fig 45 Menu Part 1 V6 Integral Process Thermostats and High Temperature Thermostats 89 198 ...
Страница 183: ...Fig 99 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 183 198 ...
Страница 184: ...Fig 100 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 184 198 ...
Страница 186: ...V6 Integral Process Thermostats and High Temperature Thermostats 186 198 ...
Страница 187: ...V6 Integral Process Thermostats and High Temperature Thermostats 187 198 ...
Страница 199: ......