n
Selecting a tolerance range that is too small can cause undesired delays.
In extreme cases, it may not be possible to continue the program. The
selected tolerance range should not be too small, especially if the control
is external. A greater tolerance has been entered for segment 5 to
guarantee adherence to the required time of 10 minutes, even with
transient responses (3).
n
A tolerance range should only be programmed for flat (slow) ramps,
if appropriate. Steep ramps that come close to the maximum possible
heating-up or cooling rates of the device may be severely delayed (4) if
the tolerance range is too small (in segment 2 here).
Fig. 64: Target/actual program process
The above graph of the edited process illustrates the possible delay of the
actual temperature (solid line) in relation to the set temperature of the
programmer (highlighted gray).
6.18.2
Set and process program
Please note:
n
If a segment time > 999:59 h is included, this time must be distributed
over several successive segments.
1.
Select the
Edit
menu item for the selected program.
You can edit the program.
Start processing
Fig. 65: Editing a program
V6
Integral Process Thermostats and High-Temperature Thermostats
112 / 198
Содержание IN 1030 T
Страница 89: ...6 3 Menu structure Fig 45 Menu Part 1 V6 Integral Process Thermostats and High Temperature Thermostats 89 198 ...
Страница 183: ...Fig 99 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 183 198 ...
Страница 184: ...Fig 100 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 184 198 ...
Страница 186: ...V6 Integral Process Thermostats and High Temperature Thermostats 186 198 ...
Страница 187: ...V6 Integral Process Thermostats and High Temperature Thermostats 187 198 ...
Страница 199: ......