9.
Enter the numerical values for byte 2, byte 3 and byte 4.
10. Once you have entered the numerical values, press the left arrow key.
You return to the
LAN configuration
menu.
11.
Scroll to the numerical values of the [Local mask] menu item and
press the Enter key.
The
Local mask
menu opens.
12.
Enter the numerical values, as described in points 7 to 9.
13.
Once you have entered the numerical values, press the left arrow key.
You return to the
LAN configuration
menu.
14. If required, also enter the numerical values for the [Gateway] and
[DNS server].
15.
Once you have entered the numerical values, press the left arrow key.
The entered numerical values of [Local IP address], [Local mask],
[Gateway] and [DNS server] are displayed.
16.
Press the [ANW.] softkey to accept the entered numerical values.
17.
Press the left cursor key to the move up one menu level and select the
PC control
menu item and press Enter.
18.
Confirm the [PC control] entry once more.
19.
Select the option [On] and confirm the entry.
The control for the control station is activated.
No settings are accepted if you leave the LAN configuration menu
without first pressing the
[ANW.]
key.
Set the
[DHCP client]
from
[off]
to
[on]
, all numerical values are
reset to 0. 0. 0. 0.
If you have set up an Ethernet communication between the control
station and the constant temperature equipment, it may take 1 or
2 minutes to establish the connection.
1.
Start the Windows command processor by entering
cmd.exe
@
on
the PC with Microsoft Windows operating system.
The entry window will open.
Checking the LAN network
V6
Integral Process Thermostats and High-Temperature Thermostats
48 / 198
Содержание IN 1030 T
Страница 89: ...6 3 Menu structure Fig 45 Menu Part 1 V6 Integral Process Thermostats and High Temperature Thermostats 89 198 ...
Страница 183: ...Fig 99 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 183 198 ...
Страница 184: ...Fig 100 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 184 198 ...
Страница 186: ...V6 Integral Process Thermostats and High Temperature Thermostats 186 198 ...
Страница 187: ...V6 Integral Process Thermostats and High Temperature Thermostats 187 198 ...
Страница 199: ......