Unit
IN 2050 PW
Pump level
IN 2560 PW
IN 2560 PW
Pump level
Recommended
cooling water tem-
perature
°C
15
---
15
15
---
Recommended
cooling water pres-
sure difference
bar
3
---
3
3
---
Minimum cooling
water pressure differ-
ence*
bar
0.8
---
0.8
0.8
---
Maximum cooling
water pressure differ-
ence
bar
10
---
10
10
---
Maximum cooling
water pressure
bar
10
---
10
10
---
* only approved for a max. cooling water temperature of 20 °C
Table 55: Cooling output
Unit
IN 280 XT
IN 280 XTW
IN 590 XTW
Pump level
Cooling output at (ambient tem-
perature 20 °C)
200 °C
kW
1.60
1.70
4.50
8
100 °C
kW
1.60
1.70
4.50
8
20 °C
kW
1.60
1.70
4.50
8
10 °C
kW
1.55
1.65
4.45
8
0 °C
kW
1.50
1.60
4.40
8
-10 °C
kW
1.50
1.60
4.40
8
-20 °C
kW
1.70
1.80
4.60
4
-30 °C
kW
1.70
1.80
4.60
4
-40 °C
kW
1.65
1.80
4.50
4
-50 °C
kW
1.40
1.50
4.20
4
-60 °C
kW
0.85
0.90
2.70
4
-70 °C
kW
0.35
0.45
1.40
4
-80 °C
kW
0.15
0.18
0.60
À
2
-90 °C
kW
---
---
0.20
2
Cooling medium
---
Air
Water
Water
---
Cooling water flow rate
l/min
---
2.7
15.8
---
Cooling water temperature
minimum/maximum
°C
---
10/30
10/30
---
Integral XT with two-stage cooling unit
V6
Integral Process Thermostats and High-Temperature Thermostats
172 / 198
Содержание IN 1030 T
Страница 89: ...6 3 Menu structure Fig 45 Menu Part 1 V6 Integral Process Thermostats and High Temperature Thermostats 89 198 ...
Страница 183: ...Fig 99 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 183 198 ...
Страница 184: ...Fig 100 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 184 198 ...
Страница 186: ...V6 Integral Process Thermostats and High Temperature Thermostats 186 198 ...
Страница 187: ...V6 Integral Process Thermostats and High Temperature Thermostats 187 198 ...
Страница 199: ......