n
Ambient temperature from 5°C to 40°C
n
Maximum relative humidity 80% at an ambient temperature of 31°C,
relative humidity linearly decreasing to 50% at 40°C.
n
Fluctuations of the mains voltage up to ±10 % of the nominal voltage.
n
Surge category II
1.6.3 Time limits
Service life
- The device is designed for 20,000 operating
hours.
Maintenance intervals
- See
Chapter 9.2 “Maintenance intervals”
Maximum operating
period without interrup-
tion
- The device is designed for supervised contin-
uous operation.
1.7
Prohibition of modifications to the device
Any technical modification of the device by the user is prohibited. Any
damage resulting from unauthorized modification is not covered by customer
service or the product warranty. Service work may only be performed by the
LAUDA Service department or a service partner authorized by LAUDA.
1.8
Materials
All parts that come into contact with heat transfer liquid are manufac-
tured from high-quality materials adapted to withstand the operating tem-
perature. High-quality stainless steels, copper, brass and premium-quality
heat-resistant plastics are used.
1.9
Fluorinated refrigerant
Refrigeration process thermostats are operated with fluorinated refrigerants.
The designation and refrigerant charge are specified on the type plate.
1.10
Requirements for the heat transfer liquid
n
Heat transfer liquids are used to control the temperature. LAUDA heat
transfer liquids are recommended for the constant temperature equip-
ment. LAUDA heat transfer liquids have been tested by the company
LAUDA DR. R. WOBSER GMBH & CO. KG and approved for this
device.
n
The heat transfer liquids are suitable for a specific temperature range.
This temperature range must correspond with the temperature range of
your application.
n
Hazards caused by high or low temperatures or fire may arise during
operation if the heat transfer liquid exceeds or falls below certain tem-
peratures or if the container ruptures causing a reaction with the heat
transfer liquid.
n
The safety data sheet of the heat transfer liquid specifies hazards and
the corresponding safety measures required for handling the liquid. The
safety data sheet of the heat transfer liquid must therefore be observed
to ensure proper use of the device.
V6
Integral Process Thermostats and High-Temperature Thermostats
12 / 198
Содержание IN 1030 T
Страница 89: ...6 3 Menu structure Fig 45 Menu Part 1 V6 Integral Process Thermostats and High Temperature Thermostats 89 198 ...
Страница 183: ...Fig 99 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 183 198 ...
Страница 184: ...Fig 100 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 184 198 ...
Страница 186: ...V6 Integral Process Thermostats and High Temperature Thermostats 186 198 ...
Страница 187: ...V6 Integral Process Thermostats and High Temperature Thermostats 187 198 ...
Страница 199: ......