CAUTION!
Risk of external consumer bursting
Scalding, cold burns
Adjust the pump pressure with the bypass.
CAUTION!
Risk of heat transfer liquid escaping from consuming unit in an
elevated location
Electric shock
If the external consuming unit is positioned above the
device, heat transfer liquid may escape from the device
when the pump stops. Therefore use the return protection
available as an accessory in the external hydraulic circuit.
The following information is only relevant for the Integral XT device:
CAUTION!
Risk of external consuming unit bursting
Scalding, cold burns
If the external consuming unit is located in a lower position
and is sensitive to pressure, also take into account the
additional pressure resulting from the difference in height
between the consuming unit and the device.
Please note:
n
Only connect hydraulically closed applications to the device!
n
Use the shortest possible hoses with the largest possible diameter in the
external circuit.
If the diameter of the hose is too small, there will be a temperature
gradient between the device and the external application due to the low
flow rate. In this case, increase the outflow temperature or pump level
accordingly.
n
Secure the hoses to the hose nozzles using hose clips.
n
If using a pressure-sensitive application (such as a glass reactor), use a
pressure relief device.
n
Open any shut-off valves in the external application. Only switch on the
device when flow is possible through the external application.
n
Depending on the type of application, a vent valve can significantly
simplify the aeration process. The vent valve must be located at the
highest point of the circuit (Fig. 19).
n
Reactors for steam heating are not suitable as external applications, as
they usually have a non-flow-through area in which gas cushions can
form.
n
If external control is used, a Pt100 sensor must be installed in the
external application or the external temperature signal is transmitted via
an interface module.
V6
Integral Process Thermostats and High-Temperature Thermostats
38 / 198
Содержание IN 1030 T
Страница 89: ...6 3 Menu structure Fig 45 Menu Part 1 V6 Integral Process Thermostats and High Temperature Thermostats 89 198 ...
Страница 183: ...Fig 99 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 183 198 ...
Страница 184: ...Fig 100 Declaration of Conformity V6 Integral Process Thermostats and High Temperature Thermostats 184 198 ...
Страница 186: ...V6 Integral Process Thermostats and High Temperature Thermostats 186 198 ...
Страница 187: ...V6 Integral Process Thermostats and High Temperature Thermostats 187 198 ...
Страница 199: ......