Screen 7 - Alarms summary
This screen will detail all alarm conditions set by the operator. It will detail the
condition for the alarm to trigger, the status (inactive, triggered, latched, or
recovering), the time and date of the alarm, and the associated relay.
Pressing key ‘4’ and key ‘6’ scrolls the user through the relays available for
alarms. This will also summarize inactive alarms.
Pressing the left soft key (‘Mute’) on an active alarm will disable the alarm and
re-energizes the associated relay until the condition is cleared.
Pressing the left soft key (‘Clear Latch’) on a latched alarm will unlatch the
alarm, clearing its status and re-energizes the associated relay.
5.6
Stop Sampling
From the ‘Gas Readings Screen’, pressing the middle soft key will ‘Stop
Sampling’. At this point, the sampling process is stopped, all solenoid valves
are closed, and the outputs are frozen at their last known value. The following
screen will be presented to the user: