Operating Manual
141
© Copyright 2016
OMBG3000 – Rev1
3
Disconnect the QRC from the top
of the catchpot.
4
Open the drain valve to allow
liquid to empty from the
catchpot.
Warning
The catchpot can hold approximately 90ml of liquid.
Dependent upon the application the liquid removed may
be contaminated and should be discharged to an area
where it is safe to do so.
This line may also vent sample gas for a brief period
during each draining operation if the sample inlet and
gas out valves are not closed.
5
Once the catchpot has emptied,
close the drain valve.