7.4
Replacing the Catchpot Filter
Warning
Do not open when an explosive atmosphere is present.
Mains voltages are present within the BIOGAS 3000
system and great care needs to be observed. If there is
any uncertainty, seek advice from a professional or
isolate the supply. Alternatively, contact our technical
support team at (800) 968-2026 or email
(Please note a charge may be applicable.)
The coalescing catchpot filter should be replaced if showing signs of
contamination. Failure to replace the filter will result in the gas flow being
restricted or blocked to the BIOGAS 3000 module. In addition, it will cause a
flow fail error.
Op. Image
Instruction
1
No image
Drain the catchpot of any
contents. Refer to
2
From the ‘Gas Readings Screen’,
press the middle soft key to ‘Stop
Sampling’.
Note: If required, press the
middle soft key once more
to fix the outputs at their
safe values.
3
Ensure the drain, gas inlet, and
gas outlet valves are closed.