protective cover will need to be fitted.
Please refer to the following table for the designation and location of each fuse
together with its type and rating:
Designation Circuit
Rated
voltage
Rated
current
Type
FS1
Mains input
Power supplies
Heater
250v
3.15A
20mm time delay,
ceramic
FS2
24Vdc pumps
Drain pump
Reed switch
250v
1.25A
20mm time delay,
ceramic
FS3
4-20mA loop
Relay control
External sensor
250v
500mA
20mm time delay,
ceramic
FS4
Solenoids
250V
1A
20mm time delay,
ceramic
FS5
24Vdc supply
250V
2A
20mm time delay,
ceramic
FS6
BIOGAS 3000
module supply
250v
1A
20mm time delay,
ceramic
FS7
12Vdc supply to
Interface PCB
250V
1A
20mm time delay,
ceramic
Note: Please contact our Technical Support team at (800) 968-2026 or
email
9.10
User Calibration Explained
General
9.10.1
User calibration is a means of optimising the performance of the BIOGAS 3000
system to the current operating conditions such as temperature and pressure
as well as correcting for analyser drift caused by the infrared source.
User calibration has two operations, zero and span, and each may be
performed individually. However, for a complete user calibration both must be
completed.