152
Operating Manual
OMBG3000 – Iss.01
© Copyright 2016
General Set-Up
7.6.3
The image below shows an example of how the equipment will look when
performing the test procedure.
Note: Due to the significant differences between models, each set-up will
be slightly different to the image shown.
Performing the Pressure Test
7.6.4
Warning
Do not open when an explosive atmosphere is present.
Ensure the gas supply has been isolated at the source
before disconnecting the tubing from the equipment.
The test is to apply 100mbar of pressure to each gas inlet, including air, in turn
as per the table associated to the model types below. Solenoids are to be
toggled via the
. A successful test is a pressure drop
of less than 10mbar in a minute on each test point.
For each test, the ball valves on the ‘Process Return’ and ‘Vent to Atmosphere’
lines should be closed.