Flow Chart 3 - Gas Check and Calibration
Start
Select gas mix
Sampling stops
and outputs are
frozen at their last
known reading
Disconnect
QRC from
catchpot
Fresh air or
gas bottle?
Connect
gas to QRC
Span O
2
only or
span O
2
and zero
others?
Press
‘Check’
Wait for
countdown
Check
‘status’ of
gases
Retry, Set
Calibration, or
Record
No
Air
Gas
Retry
Records gas check
to the event log.
Does not calibrate
system
Record
Was calibration
successful?
Calibrate
Updates
calibration of
system
Yes
Refer to
‘Problem
Solving’ section
No
Fresh air or
gas bottle?
Turn off
gas and
disconnect
Reconnect
QRC
Gas
Air
No
Set calibration
valve to ‘span’
Press
‘Reconnect’
Gas concentration
greater than ext.
cell range?
Yes
Wait for air
purge
Does gas mix need
updating to match
bottle?
Edit gas
mix
Yes
Turn gas in
and gas out
valves off
Set calibration
valve to ‘OFF’
Turn gas in
and gas out
valves on