154
Operating Manual
OMBG3000 – Iss.01
© Copyright 2016
7.6.4.2
BG3K2
Apply
pressure to:
Solenoid
1
Solenoid
2
Solenoid
5
Solenoid
6
Air Purge
OFF
OFF
ON
OFF
Sample Gas 1
OFF
OFF
OFF
OFF
Sample Gas 2
ON
OFF
OFF
ON
Note: Ensure the ball valves on the ‘Process Return’ and ‘Vent to
Atmosphere’ lines are opened once testing is completed.
Note: When reattaching the sample pipes to the system, ensure all tube
connections are tight and free from leaks.
Note: If assistance is required, please contact our technical support team
at (800) 968-2026 or email