Lake Shore Model 325 Temperature Controller User’s Manual
Remote Operation
6-1
CHAPTER 6
COMPUTER INTERFACE OPERATION
6.0
GENERAL
This chapter provides operational instructions for the computer interface for the Lake Shore Model 325 temperature
controller. Either of the two computer interfaces provided with the Model 325 permit remote operation. The first is the
IEEE-488 interface described in Section 6.1. The second is the serial interface described in Section 6.2. The two
interfaces share a common set of commands detailed in Section 6.3. Only one of the interfaces can be used at a time.
6.1
IEEE-488 INTERFACE
The IEEE-488 interface is an instrumentation bus with hardware and programming standards that simplify instrument
interfacing. The Model 325 IEEE-488 interface complies with the IEEE-488.2-1987 standard and incorporates its
functional, electrical, and mechanical specifications unless otherwise specified in this manual.
All instruments on the interface bus perform one or more of the interface functions of TALKER, LISTENER, or BUS
CONTROLLER. A TALKER transmits data onto the bus to other devices. A LISTENER receives data from other
devices through the bus. The BUS CONTROLLER designates to the devices on the bus which function to perform.
The Model 325 performs the functions of TALKER and LISTENER but cannot be a BUS CONTROLLER. The BUS
CONTROLLER is the digital computer that tells the Model 325 which functions to perform.
Below are Model 325 IEEE-488 interface capabilities:
•
SH1:
Source handshake capability.
•
RL1:
Complete remote/local capability.
•
DC1:
Full device clear capability.
•
DT0:
No device trigger capability.
•
C0:
No system controller capability.
•
T5:
Basic TALKER, serial poll capability, talk only, unaddressed to talk if addressed to listen.
•
L4:
Basic LISTENER, unaddressed to listen if addressed to talk.
•
SR1:
Service request capability.
•
AH1:
Acceptor handshake capability.
•
PP0:
No parallel poll capability.
•
E1:
Open collector electronics.
Instruments are connected to the IEEE-488 bus by a 24-conductor connector cable as specified by the standard. Refer to
Section 8.7.2. Cables can be purchased from Lake Shore or other electronic suppliers.
Cable lengths are limited to 2 m for each device and 20 m for the entire bus. The Model 325 can drive a bus with up to
10 loads. If more instruments or cable length is required, a bus expander must be used.
6.1.1
Changing IEEE-488 Interface Parameters
Two interface parameters, address and terminators, must be set from the front panel before communication with the
instrument can be established. Other interface parameters can be set with device specific commands using the interface
(Section 6.3).
Press the
Interface
key. The first screen is for selecting the serial interface baud rate, and can be skipped by pressing the
Enter
key. The address screen is then displayed as follows.
Select-With-°®
IEEE-Address--12
Содержание 325
Страница 4: ...Lake Shore Model 325 Temperature Controller User s Manual...
Страница 6: ......
Страница 22: ...Lake Shore Model 325 Temperature Controller User s Manual 1 10 Introduction This Page Intentionally Left Blank...
Страница 47: ...Lake Shore Model 325 Temperature Controller User s Manual 3 11 This Page Intentionally Left Blank...
Страница 48: ...Lake Shore Model 325 Temperature Controller User s Manual 3 12 This Page Intentionally Left Blank...
Страница 76: ...Lake Shore Model 325 Temperature Controller User s Manual 4 28 Operation This Page Intentionally Left Blank...
Страница 122: ...Lake Shore Model 325 Temperature Controller User s Manual 6 36 Remote Operation This Page Intentionally Left Blank...
Страница 128: ...Lake Shore Model 325 Temperature Controller User s Manual 7 6 Options and Accessories...