Lake Shore Model 325 Temperature Controller User’s Manual
Remote Operation
6-31
MOUT
Control Loop Manual Heater Power (MHP) Output Command
Input:
MOUT <loop>, <value>[term]
Format:
n,
nnnnnn[term]
<loop>
Specifies loop to configure: 1 or 2.
<value>
Specifies value for manual output.
Example:
MOUT 1,22.45[term]
– Control Loop 1 manual heater output is 22.45%.
MOUT?
Control Loop Manual Heater Power (MHP) Output Query
Input:
MOUT? <loop>[term]
Format:
n
<loop>
Specifies which loop to query: 1 or 2.
Returned:
<value>
Format:
nnnnnn[term]
(Refer to command for description)
PID
Control Loop PID Values Command
Input:
PID <loop>, <P value>, <I value>, <D value>[term]
Format:
n,±nnnnnn,±nnnnnn,±nnnnnn
<loop>
Specifies loop to configure: 1 or 2.
<P value>
The value for control loop Proportional (gain): 0.1 to 1000.
<I value>
The value for control loop Integral (reset): 0.1 to 1000.
<D value>
The value for control loop Derivative (rate): 0 to 200.
Remarks:
Setting resolution is less than 6 digits indicated.
Example:
PID 1,10,50[term]
– Control Loop 1 P is 10 and I is 50.
PID?
Control Loop PID Values Query
Input:
PID? <loop>[term]
Format:
n
<loop>
Specifies which loop to query: 1 or 2.
Returned:
<P value>, <I value>, <D value>[term]
Format:
±nnnnnn,±nnnnnn,±nnnnnn
(Refer to command for description)
RAMP
Control Setpoint Ramp Parameter Command
Input:
RAMP <loop>, <off/on>, <rate value>[term]
Format:
n,n,±nnnnn
<loop>
Specifies which loop to configure: 1 or 2.
<off/on>
Specifies whether ramping is 0 = Off or 1 = On.
<rate value>
Specifies setpoint ramp rate in kelvin per minute from 0.0 to 100. The rate is always
positive, but will respond to ramps up or down. A ramp setting of 0.0 will cause the
instrument to respond as if the ramp is off, i.e. setpoint changes will be immediate.
Example:
RAMP 1,1,10.5[term]
– When Control Loop 1 setpoint is changed, ramp the current setpoint to the
target setpoint at 10.5 K/minute.
Содержание 325
Страница 4: ...Lake Shore Model 325 Temperature Controller User s Manual...
Страница 6: ......
Страница 22: ...Lake Shore Model 325 Temperature Controller User s Manual 1 10 Introduction This Page Intentionally Left Blank...
Страница 47: ...Lake Shore Model 325 Temperature Controller User s Manual 3 11 This Page Intentionally Left Blank...
Страница 48: ...Lake Shore Model 325 Temperature Controller User s Manual 3 12 This Page Intentionally Left Blank...
Страница 76: ...Lake Shore Model 325 Temperature Controller User s Manual 4 28 Operation This Page Intentionally Left Blank...
Страница 122: ...Lake Shore Model 325 Temperature Controller User s Manual 6 36 Remote Operation This Page Intentionally Left Blank...
Страница 128: ...Lake Shore Model 325 Temperature Controller User s Manual 7 6 Options and Accessories...