Lake Shore Model 325 Temperature Controller User’s Manual
Operation
4-15
4.7
CONTROL SETUP
After the Input Setup has been completed (Section 4.4) and Loop is selected (Section 4.6.1), the user can begin to setup
temperature control parameters.
Control input is the sensor input that is used for control feedback in closed loop control. Either Input A or B can be
assigned to either Loop 1 or 2. It is not recommended to assign both loops to one input. Control input is ignored when
using open loop control mode. To change control input, press the
Control Setup
key and the following screen will
appear.
Select for Loop 1 °®
Control with Input A
Use the
s
or
t
key to toggle between Input A and B. Press the
Enter
key to accept the setting and continue with
additional selections. You can press the
Escape
key any time to exit the routine.
The control setpoint can be displayed and set in temperature or sensor units. Changing setpoint units does not change
operation of the controller, only the way the setpoint is displayed and entered. A valid curve must be assigned to the
control input to use temperature units. To change setpoint units press the
Control Setup
key and press
Enter
until the
following display appears.
Select for Loop 1 °®
SP Units Temp K
Use the
s
or
t
key to cycle through the following setpoint units: Temp K, Temp C, and Sensor, where K = kelvin,
C = degrees Celsius, and Sensor = volts (V), millivolts (mV), or ohms (
). Press the
Enter
key.
The Model 325 has two control modes, Closed Loop and Open Loop. Closed Loop control, often called feedback
control, is described in Section 2.6 of this manual. During closed loop control operation, the Control Input, Setpoint,
Heater Range, PID, and Manual Heater Power (MHP) output parameters are active. Open loop control mode allows the
user to directly set the heater output for Loop 1 or Loop 2 with the MHP Output parameter. During Open Loop control
only the heater range and MHP Output parameters are active. To change Control Mode press the
Control Setup
key and
press
Enter
until the following display appears.
Select for Loop 1 °®
Control Mode Closed
Use the
s
or
t
key to toggle between open and closed loop control. Press the
Enter
key.
The Power Up setting refers to how the control output will respond after the instrument is powered down. Power Up
Enable means the controller will power up with the control output in the same state it was before power was turned off.
Power Up Disable means the controller will always power up with the heater off no matter how it was set when power
was turned off. To change the Power Up parameter press the Control Setup key and press Enter until the following
display appears.
Select for Loop 1 °®
Power Up Disable
Use the
s
or
t
key to toggle between Power Up Enable and Disable. Press the
Enter
key.
Содержание 325
Страница 4: ...Lake Shore Model 325 Temperature Controller User s Manual...
Страница 6: ......
Страница 22: ...Lake Shore Model 325 Temperature Controller User s Manual 1 10 Introduction This Page Intentionally Left Blank...
Страница 47: ...Lake Shore Model 325 Temperature Controller User s Manual 3 11 This Page Intentionally Left Blank...
Страница 48: ...Lake Shore Model 325 Temperature Controller User s Manual 3 12 This Page Intentionally Left Blank...
Страница 76: ...Lake Shore Model 325 Temperature Controller User s Manual 4 28 Operation This Page Intentionally Left Blank...
Страница 122: ...Lake Shore Model 325 Temperature Controller User s Manual 6 36 Remote Operation This Page Intentionally Left Blank...
Страница 128: ...Lake Shore Model 325 Temperature Controller User s Manual 7 6 Options and Accessories...