Lake Shore Model 325 Temperature Controller User’s Manual
Remote Operation
6-27
CSET
Control Loop Parameter Command
Input:
CSET <loop>,
<input>,
<units>,
<powerup enable>,
<current/power>[term]
Format:
n,a,n,n,n
<loop>
Specifies which loop to configure: 1 or 2.
<input>
Specifies which input to control from: A or B.
<units>
Specifies setpoint units. Valid entries: 1 = kelvin, 2 = Celsius,
3 = sensor units.
<powerup enable>
Specifies whether the control loop is on or off after power-up, where
0 = powerup enable off and 1 = powerup enable on.
<current/power>
Specifies whether the heater output displays in current or power. Valid entries:
1 = current or 2 = power.
Example:
CSET 1,A,1,1[term]
– Control Loop 1 controls off of Input A with setpoint in kelvin.
CSET?
Control Loop Parameter Query
Input:
CSET? <loop>[term]
Format:
n
<loop>
Specifies which loop to query: 1 or 2.
Returned:
<input>, <units>, <powerup enable>, <current/power>[term]
Format:
a,n,n,n
(Refer to command for description)
DFLT
Factory Defaults Command
Input:
DFLT 99[term]
Remarks:
Sets all configuration values to factory defaults and resets the instrument. The "99" is included to
prevent accidentally setting the unit to defaults.
DISPFLD
Displayed Field Command
Input:
DISPFLD <field>, <item>, <source>[term]
Format:
n,n,n
<field>
Specifies field to configure: 1
–
4.
<item>
Specifies item to display in the field: 0 = Off, 1 = Input A, 2 = Input B,
3 = Setpoint, 4 = Heater Output.
<source>
If Item is 1 or 2, specifies input data to display. Valid entries: 1 = kelvin,
2 = Celsius, 3 = sensor units
Example:
DISPFLD 2,1,1[term]
– Displays kelvin reading for Input A in display field 2.
DISPFLD?
Displayed Field Query
Input:
DISPFLD? <field>[term]
Format:
n
<field>
Specifies field to query: 1
–
4.
Returned:
<item>, <source>[term]
Format:
n,n
(Refer to command for description)
Содержание 325
Страница 4: ...Lake Shore Model 325 Temperature Controller User s Manual...
Страница 6: ......
Страница 22: ...Lake Shore Model 325 Temperature Controller User s Manual 1 10 Introduction This Page Intentionally Left Blank...
Страница 47: ...Lake Shore Model 325 Temperature Controller User s Manual 3 11 This Page Intentionally Left Blank...
Страница 48: ...Lake Shore Model 325 Temperature Controller User s Manual 3 12 This Page Intentionally Left Blank...
Страница 76: ...Lake Shore Model 325 Temperature Controller User s Manual 4 28 Operation This Page Intentionally Left Blank...
Страница 122: ...Lake Shore Model 325 Temperature Controller User s Manual 6 36 Remote Operation This Page Intentionally Left Blank...
Страница 128: ...Lake Shore Model 325 Temperature Controller User s Manual 7 6 Options and Accessories...