Lake Shore Model 325 Temperature Controller User’s Manual
Service
8-15
Table 8-2. Calibration Table for Resistive Ranges
Range
Calibration Resistor
Nominal Value
Resistor Value
Known to
Reversal
Cal. Command
Type Number
100
Plat/250
100
±0.0070
Off
2
100
Plat/250
100
±0.0070
On
10
100
Plat/500
500
±0.0270
Off
3
100
Plat/500
500
±0.0270
On
11
1000
Plat
5 k
±0.520
Off
4
1000
Plat
5 k
±0.520
On
12
NTC RTD
5 k
±0.520
Off
5
NTC RTD
5 k
±0.520
On
13
8.12.3 Diode Sensor Input Calibration
– 1 mA Excitation Current
The instrument uses the calibration constants determined in the Diode Input Ranges Calibration (Section 8.12.2.3), for
the diode ranges that use 1 mA excitation. Therefore, no additional calibration is necessary.
NOTE:
Standard diode curves and typical sensor performance (Table 1-2) are calculated using 10 µA
excitation. Sensor temperature response characteristics will be altered if 1 mA excitation is selected.
8.12.4 Thermocouple Sensor Input Calibration
Overview
Each thermocouple sensor input requires calibration. The sensor inputs contain multiple gain stages to accommodate the
various thermocouple sensors the Model 325 supports. The input circuitry is not adjusted during calibration. Instead,
precision voltages are supplied to each input and mathematical calibration constants are calculated and programmed into
the Model 325. Constants are stored to compensate for both input offset and gain errors. Thermocouple inputs do not use
the current source.
Calibration Process
8.9.4.1
Sensor Input Calibration Setup
Allow the Model 325 to warm up for at least 1 hour with shorts placed across all thermocouple sensor inputs. If
calibrating a dual thermocouple Model 325, leave a short across the input not currently being calibrated. If the other
input is diode/resistor, place a 100 k
resistor on the input.
CAUTION:
All thermocouple connections must be tight and direct with no unnecessary jumpers or connections.
8.12.4.2
Thermocouple Input Ranges Calibration
Purpose
To determine the input offset and gain errors when the input is configured for the thermocouple ranges and provide
offset and gain calibration constants back to the Model 325.
Process
1. Configure the input for the thermocouple range to be calibrated. Turn Room Cal off.
2. Reset the calibration constants to their default values using the
CALZ
and
CALG
commands.
EXAMPLE:
Input: A
Range: Thermo/25mV
Zero Offset Command:
CALZ A,6,0
Gain Command:
CALG A,6,1
3. Short the V+ and V– terminals together, do not tie the terminals to ground.
4. Via the interface obtain the input reading using the
CALREAD?
command and record this number.
Содержание 325
Страница 4: ...Lake Shore Model 325 Temperature Controller User s Manual...
Страница 6: ......
Страница 22: ...Lake Shore Model 325 Temperature Controller User s Manual 1 10 Introduction This Page Intentionally Left Blank...
Страница 47: ...Lake Shore Model 325 Temperature Controller User s Manual 3 11 This Page Intentionally Left Blank...
Страница 48: ...Lake Shore Model 325 Temperature Controller User s Manual 3 12 This Page Intentionally Left Blank...
Страница 76: ...Lake Shore Model 325 Temperature Controller User s Manual 4 28 Operation This Page Intentionally Left Blank...
Страница 122: ...Lake Shore Model 325 Temperature Controller User s Manual 6 36 Remote Operation This Page Intentionally Left Blank...
Страница 128: ...Lake Shore Model 325 Temperature Controller User s Manual 7 6 Options and Accessories...