Lake Shore Model 325 Temperature Controller User’s Manual
1-4
Introduction
1.1
SENSOR SELECTION
Table 1-1. Sensor Temperature Range
Model
Useful Range
Magnetic Field Use
Diodes
Silicon Diode
DT-670-SD
1.4 K to 500 K
T
60 K & B
3 T
Silicon Diode
DT-670E-BR
30 K to 500 K
T
60 K & B
3 T
Silicon Diode
DT-414
1.4 K to 375 K
T
60 K & B
3 T
Silicon Diode
DT-421
1.4 K to 325 K
T
60 K & B
3 T
Silicon Diode
DT-470-SD
1.4 K to 500 K
T
60 K & B
3 T
Silicon Diode
DT-471-SD
10 K to 500 K
T
60 K & B
3 T
GaAlAs Diode
TG-120-P
1.4 K to 325 K
T
4.2 K & B
5 T
GaAlAs Diode
TG-120-PL
1.4 K to 325 K
T
4.2 K & B
5 T
GaAlAs Diode
TG-120-SD
1.4 K to 500 K
T
4.2 K & B
5 T
Positive Temperature
Coefficient (PTC) RTDs
100
Platinum
PT-102/3
14 K to 873 K
T > 40 K & B
2.5 T
100
Platinum
PT-111
14 K to 673 K
T > 40 K & B
2.5 T
Rhodium-Iron
RF-800-4
1.4 K to 500 K
T > 77 K & B
8 T
Rhodium-Iron
RF-100T/U
1.4 K to 325 K
T > 77 K & B
8 T
Negative Temperature
Coefficient (NTC) RTDs
1
Cernox™
CX-1010
2 K to 325 K
4
T > 2 & B
19 T
Cernox
CX-1030-HT
3.5 K to 420 K
2,5
T > 2 & B
19 T
Cernox
CX-1050-HT
4 K to 420 K
2,5
T > 2 & B
19 T
Cernox
CX-1070-HT
15 K to 420 K
2
T > 2 & B
19 T
Cernox
CX-1080-HT
50 K to 420 K
2
T > 2 & B
19 T
Germanium
GR-200A/B-1000
2.2 K to 100 K
3
Not Recommended
Germanium
GR-200A/B-1500
2.6 K to 100 K
3
Not Recommended
Germanium
GR-200A/B-2500
3.1 K to 100 K
3
Not Recommended
Carbon-Glass
CGR-1-500
4 K to 325 K
4
T > 2 K to
19 T
Carbon-Glass
CGR-1-1000
5 K to 325 K
4
T > 2 K to
19 T
Carbon-Glass
CGR-1-2000
6 K to 325 K
4
T > 2 K to
19 T
Rox™
RX-102A
1.4 K to 40 K
4
T > 2 K to
10 T
Thermocouples
Type K
9006-006
3.2 K to 1505 K
Not Recommended
Type E
9006-004
3.2 K to 934 K
Not Recommended
Chromel-AuFe 0.07%
9006-002
1.2 K to 610 K
Not Recommended
1
Single excitation current may limit the low temperature range of NTC resistors.
2
Non-HT version maximum temperature: 325 K.
3
Low temperature limited by input resistance range.
4
Low temperature specified with self-heating error:
5 mK.
5
Low temperature specified with self-heating error:
12 mK.
Silicon diodes
are the best choice for general cryogenic use from 1.4 K to above room temperature. Diodes are
economical to use because they follow a standard curve and are interchangeable in many applications. They are
not suitable for use in ionizing radiation or magnetic fields.
Cernox™
thin-film RTDs offer high sensitivity and low magnetic field-induced errors over the 2 K to 420 K
temperature range. Cernox sensors require calibration.
Platinum RTDs
offer high uniform sensitivity from 30 K to over 800 K. With excellent reproducibility, they are
useful as thermometry standards. They follow a standard curve above 70 K and are interchangeable in many
applications.
Содержание 325
Страница 4: ...Lake Shore Model 325 Temperature Controller User s Manual...
Страница 6: ......
Страница 22: ...Lake Shore Model 325 Temperature Controller User s Manual 1 10 Introduction This Page Intentionally Left Blank...
Страница 47: ...Lake Shore Model 325 Temperature Controller User s Manual 3 11 This Page Intentionally Left Blank...
Страница 48: ...Lake Shore Model 325 Temperature Controller User s Manual 3 12 This Page Intentionally Left Blank...
Страница 76: ...Lake Shore Model 325 Temperature Controller User s Manual 4 28 Operation This Page Intentionally Left Blank...
Страница 122: ...Lake Shore Model 325 Temperature Controller User s Manual 6 36 Remote Operation This Page Intentionally Left Blank...
Страница 128: ...Lake Shore Model 325 Temperature Controller User s Manual 7 6 Options and Accessories...