Lake Shore Model 325 Temperature Controller User’s Manual
Remote Operation
6-29
IEEE?
IEEE-488 Interface Parameter Query
Input:
IEEE?[term]
Returned:
<terminator>, <EOI enable>, <address>[term]
Format:
n,n,nn
(Refer to command for description)
INCRV
Input Curve Number Command
Input:
INCRV <input>, <curve number>[term]
Format:
a,nn
<input>
Specifies which input to configure: A or B.
<curve number>
Specifies which curve the input uses. If specified curve parameters do not match
the input, the curve number defaults to 0. Valid entries:
0 = none, 1
–
20 = standard curves, 21
–
35 = user curves.
Remarks:
Specifies the curve an input uses for temperature conversion.
Example:
INCRV A,23[term]
– Input A uses User Curve 23 for temperature conversion.
INCRV?
Input Curve Number Query
Input:
INCRV? <input>[term]
Format:
a
<input>
Specifies which input to query: A or B.
Returned:
<curve number>[term]
Format:
nn
(Refer to command for description)
INTYPE
Input Type Parameter Command
Input:
INTYPE <input>, <sensor type>, <compensation>[term]
Format:
a,n,n
<input>
Specifies input to configure: A or B.
<sensor type>
Specifies input sensor type. Valid entries:
0 = Silicon diode
5 = NTC RTD
1 = GaAlAs diode
6 = Thermocouple 25 mV
2 = 100
platinum/250
7 = Thermocouple 50 mV
3 = 100
platinum/500
8 = 2.5 V, 1 mA
4 = 1000
platinum
9 = 7.5 V, 1 mA
<compensation>
Specifies input compensation where 0 = off and 1 = on. Reversal for thermal EMF
compensation if input is resistive, room temperature compensation if input is
thermocouple. Always 0 if input is a diode.
Example:
INTYPE A,0,0[term]
– Sets Input A sensor type to silicon diode.
INTYPE?
Input Type Parameter Query
Input:
INTYPE? <input>[term]
Format:
a
<input>
Specifies input to query: A or B.
Returned:
<sensor type>, <compensation>[term]
Format:
n,n
(Refer to command for description)
Содержание 325
Страница 4: ...Lake Shore Model 325 Temperature Controller User s Manual...
Страница 6: ......
Страница 22: ...Lake Shore Model 325 Temperature Controller User s Manual 1 10 Introduction This Page Intentionally Left Blank...
Страница 47: ...Lake Shore Model 325 Temperature Controller User s Manual 3 11 This Page Intentionally Left Blank...
Страница 48: ...Lake Shore Model 325 Temperature Controller User s Manual 3 12 This Page Intentionally Left Blank...
Страница 76: ...Lake Shore Model 325 Temperature Controller User s Manual 4 28 Operation This Page Intentionally Left Blank...
Страница 122: ...Lake Shore Model 325 Temperature Controller User s Manual 6 36 Remote Operation This Page Intentionally Left Blank...
Страница 128: ...Lake Shore Model 325 Temperature Controller User s Manual 7 6 Options and Accessories...