Lake Shore Model 325 Temperature Controller User’s Manual
1-6
Introduction
1.2
SPECIFICATIONS
Input Specifications
Sensor
Temperature
Coefficient
Input
Range
Excitation
Current
Display
Resolution
Measurement
Resolution
Electronic
Accuracy
(at 25 °C)
Measurement
Temperature
Coefficient
Electronic
Control
Stability
1
Diode
Negative
0 V to 2.5 V
10 µA ±0.05%
2,3
100 µV
0.4 µV
±80 µV
±0.005% of rdg
(10 µV +
0.0005% of rdg)/°C
±20 µV
Negative
0 V to 7.5 V
10 µA ±0.05%
2,3
100 µV
10 µV
±320 µV
±0.01% of rdg
(20 µV +
0.0005% of rdg)/°C
±40 µV
PTC RTD
Positive
0
to 500
1 mA
4
10 m
2 m
±0.004
±0.01% of rdg
(0.2 m
+
0.0005% of rdg)/°C
±4 m
Positive
0
to 5000
1 mA
4
100 m
20 m
±0.04
±0.02% of rdg
(0.2 m
+
0.0005% of rdg)/°C
±40 m
NTC RTD
Negative
0
to 7500
10 µA ±0.05%
100 m
40 m
±0.1
±0.04% of rdg
(2 m
+
0.0005% of rdg)/°C
±80 m
Thermocouple
Positive
±25 mV
NA
1 µV
0.4 µV
±1 µV
±0.05% of rdg
(0.2 µV +
0.0015% of rdg)/°C
±0.8 µV
Positive
±50 mV
NA
1 µV
20 µV
±1 µV
±0.05% of rdg
(0.2 µV +
0.0015% of rdg)/°C
±0.8 µV
1
Control stability of the electronics only, in ideal thermal system
2
Current source error has negligible effect on measurement accuracy
3
Diode input excitation can be set to 1 mA
4
Current source error is removed during calibration
5
Accuracy specification does not include errors from room temperature compensation
Thermometry
Number of inputs
2
Input configuration
Each input is factory configured for either diode / RTD or thermocouple
Isolation
Sensor inputs optically isolated from other circuits but not each other
A/D resolution
24-bit
Input accuracy
Sensor dependent, refer to Input Specifications table
Measurement resolution
Sensor dependent, refer to Input Specifications table
Max update rate:
10 rdg/s on each input, (except 5 rdg/s on input A when configured as thermocouple)
User curves
Room for 15, 200 point CalCurves
™
or user curves
SoftCal
™
Improves accuracy of DT-470 diode to ±0.25 K from 30 K to 375 K. Improves accuracy
of platinum RTDs to ±0.25 K from 70 K to 325 K. Stored as user curves.
Filter
Averages 2 to 64 input readings
Sensor Input Configuration
Diode/RTD
Thermocouple
Measurement type
4-lead differential
2-lead, room temperature compensated
Excitation
Constant current with current reversal for
RTDs
NA
Supported sensors
Diodes: Silicon, GaAlAs RTDs: 100
Platinum, 1000
Platinum, Germanium,
Carbon-Glass, Cernox, and Rox
Most thermocouple types
Standard curves
DT-470, DT-500D. DT-670, PT-100, PT-1000,
RX-102A, RX-202A
Type E, Type K, Type T, AuFe 0.07% vs. Cr,
AuFe 0.03% vs Cr
Input connector
6-pin DIN
Ceramic isothermal block
Содержание 325
Страница 4: ...Lake Shore Model 325 Temperature Controller User s Manual...
Страница 6: ......
Страница 22: ...Lake Shore Model 325 Temperature Controller User s Manual 1 10 Introduction This Page Intentionally Left Blank...
Страница 47: ...Lake Shore Model 325 Temperature Controller User s Manual 3 11 This Page Intentionally Left Blank...
Страница 48: ...Lake Shore Model 325 Temperature Controller User s Manual 3 12 This Page Intentionally Left Blank...
Страница 76: ...Lake Shore Model 325 Temperature Controller User s Manual 4 28 Operation This Page Intentionally Left Blank...
Страница 122: ...Lake Shore Model 325 Temperature Controller User s Manual 6 36 Remote Operation This Page Intentionally Left Blank...
Страница 128: ...Lake Shore Model 325 Temperature Controller User s Manual 7 6 Options and Accessories...