Lake Shore Model 325 Temperature Controller User’s Manual
6-34
Remote Operation
TEMP?
Thermocouple Junction Temperature Query
Input:
TEMP?
Returned:
<junction temperature>[term]
Format:
±nnnnnnn
Remarks:
Temperature is in kelvin. This query returns the temperature of the ceramic thermocouple block used
in the room temperature compensation calculation.
TLIMIT
Temperature Limit Command (Firmware Version 1.5 and Later)
Input:
TLIMIT <input>, <limit>[term]
Format:
a,+nnnn
<input> Specifies which input to configure: A or B.
<limit value> The temperature limit in Kelvin for which to shut down all control outputs when
exceeded. A temperature limit of zero turns the temperature limit feature off for the
given sensor input.
Remarks:
A temperature limit setting of 0 K turns the temperature limit feature off.
Example: TLIMIT B,450[term]
– If the temperature of the sensor on Input B exceeds 450 K, all control outputs
will be turned off.
TLIMIT?
Temperature Limit Query (Firmware Version 1.5 and Later)
Input:
TLIMIT? <input>[term]
Format:
a
<input> Specifies which input to query: A or B.
Returned:
<limit value>[term]
Format:
+nnnn
TUNEST?
Control Tuning Status Query
Input:
TUNEST?
Returned:
<tuning status>[term]
Format:
n
0 = no active tuning, 1 = active tuning.
Remarks:
The tuning status will return active (1) if either Loop 1 or Loop 2 is actively tuning.
ZONE
Control Loop Zone Table Parameter Command
Input:
ZONE <loop>, <zone>, <setpoint limit>, <P value>, <I value>,
<D value>, <mout value>, <range>[term]
Format:
n,nn,±nnnnnn,±nnnnnn,±nnnnnn,±nnnnnn,±nnnnnn,n[term]
<loop>
Specifies which loop to configure: 1 or 2.
<zone>
Specifies which zone in the table to configure. Valid entries are: 1
–
10.
<setpoint limit> Specifies the setpoint limit of this zone.
<P value>
Specifies the P for this zone: 0.1 to 1000.
<I value>
Specifies the I for this zone: 0.1 to 1000.
<D value>
Specifies the D for this zone: 0 to 200%.
<mout value>
Specifies the manual output for this zone: 0 to 100%.
<range>
Specifies the heater range for this zone if <loop> = 1. Valid entries: 0
–
2.
If <loop> = 2, then <range> = 1 and cannot be changed
Remarks:
Configures the control loop zone parameters. Refer to Section 2.9.
Example:
ZONE 1,1,25.0,10,20,0,0,2[term]
– Control Loop 1 zone 1 is valid to 25.0 K with P = 10,
I = 20, D = 0, and a heater range of 2.
Содержание 325
Страница 4: ...Lake Shore Model 325 Temperature Controller User s Manual...
Страница 6: ......
Страница 22: ...Lake Shore Model 325 Temperature Controller User s Manual 1 10 Introduction This Page Intentionally Left Blank...
Страница 47: ...Lake Shore Model 325 Temperature Controller User s Manual 3 11 This Page Intentionally Left Blank...
Страница 48: ...Lake Shore Model 325 Temperature Controller User s Manual 3 12 This Page Intentionally Left Blank...
Страница 76: ...Lake Shore Model 325 Temperature Controller User s Manual 4 28 Operation This Page Intentionally Left Blank...
Страница 122: ...Lake Shore Model 325 Temperature Controller User s Manual 6 36 Remote Operation This Page Intentionally Left Blank...
Страница 128: ...Lake Shore Model 325 Temperature Controller User s Manual 7 6 Options and Accessories...