Lake Shore Model 325 Temperature Controller User’s Manual
4-10
Operation
Room-Temperature Calibration Procedure (Continued)
7. The current temperature reading is displayed in kelvin.
Select for InputA
Temp 294.15½
Enter the true temperature that the thermocouple should read. If input is shorted, then enter the actual room
temperature measured by the thermometer. Press the
Enter
key to save the value.
8. To verify calibration, check that the temperature reading for the calibrated input matches the room-temperature
calibration setting value.
4.4.5
Temperature Limit (Firmware Version 1.5 and Later)
The temperature limit feature provides a means of protecting your equipment from damage by shutting down all control
outputs when the assigned temperature limit is exceeded on any sensor input. The feature is available for both of the
sensor inputs. A temperature limit of 0 K (default value) turns this feature off. This feature was added in firmware
version 1.5. If you need a firmware update, contact service (section 8.1). Section 4.18 provides the procedure to check
your firmware version.
To configure a temperature limit press the
Input Setup
key. The first screen appears as follows.
Select With °®
Input Setup
Input A
Use the
s
or
t
key to toggle between Input A and B. Press the
Enter
key until the following display appears.
Enter for Input A
Temp Limit
0K
The temperature limit is entered using the numeric keypad, which includes the numbers 0 – 9. The user can set
temperature limit in kelvin with a range of 0 to 2999 and a resolution of 1. The feature is turned off when a temperature
limit of 0 K is entered. Press the
Enter
key to save the value.
4.5
CURVE SELECTION
The Model 325 supports a variety of temperature sensors sold by Lake Shore and other manufacturers. After the
appropriate sensor type is selected for each of the two inputs (Section 4.4), an appropriate curve may be selected for
each input. The 325 can use curves from several sources. Standard curves are included with every instrument and
numbered 1
–
20. User curves, numbered 21
–
35, are loaded when a sensor does not match a standard curve. CalCurve
options are stored as user curves. SoftCal calibrations are stored as user curves or user can enter their own curves from
the front panel (Section 5.2) or computer interface (Chapter 6). The complete list of sensor curves built in to the
Model 325 is provided in Table 4-2.
During normal operation, only the curves related to the input type you have selected are displayed. If the curve you wish
to select does not appear in the selection sequence make sure the curve format matches the recommended format for the
input type selected. Refer to Table 4-1.
NOTE
:
The sensor reading of the instrument can always be displayed in sensor units. If a temperature
response curve is selected for an input, its readings may also be displayed in temperature.
Содержание 325
Страница 4: ...Lake Shore Model 325 Temperature Controller User s Manual...
Страница 6: ......
Страница 22: ...Lake Shore Model 325 Temperature Controller User s Manual 1 10 Introduction This Page Intentionally Left Blank...
Страница 47: ...Lake Shore Model 325 Temperature Controller User s Manual 3 11 This Page Intentionally Left Blank...
Страница 48: ...Lake Shore Model 325 Temperature Controller User s Manual 3 12 This Page Intentionally Left Blank...
Страница 76: ...Lake Shore Model 325 Temperature Controller User s Manual 4 28 Operation This Page Intentionally Left Blank...
Страница 122: ...Lake Shore Model 325 Temperature Controller User s Manual 6 36 Remote Operation This Page Intentionally Left Blank...
Страница 128: ...Lake Shore Model 325 Temperature Controller User s Manual 7 6 Options and Accessories...