Lake Shore Model 325 Temperature Controller User’s Manual
5-2
Advanced Operation
Table 5-1. Curve Header Parameters
Name:
The curve name cannot be changed from the front panel. Curve names can only be entered over the
computer interface (up to 15 characters). The default curve name is User xx, where xx is the curve
number.
Serial Num:
Identify specific sensors with serial numbers of up to 10 characters. The serial number field accepts
both numbers and letters, but the instrument front panel enters only numbers. To enter both numbers
and letters, enter curves over computer interface. The default is blank.
Curve
Format:
The instrument must know the data format of the curve breakpoints. Different sensor types use
different data formats. The sensor inputs require one of the formats below. The range and resolution
specified are not always available at the same time. Practical range and resolution depend on the
sensor type.
Sensor Units
Sensor Units
Format
Description
Full Scale Range
Maximum Resolution
V/K
Volts vs. kelvin
10 (V)
0.00001 (V)
/K
Resistance vs. kelvin
10 K (
)
0.001 (
)
log
/K
Log resistance vs. kelvin
4 (log
)
0.00001 (log
)
mV/K
Millivolts vs. kelvin
±100 (mV)
0.0001 (mV)
SP Limit
A setpoint temperature limit can be included with every curve. When controlling in temperature, the
setpoint cannot exceed the limit entered with the curve for the control sensor. The default is 375 K.
Set to 9999 K if no limit is required.
Coeff:
The instrument derives the temperature coefficient from the first two breakpoints. If it is set
improperly, check the first two breakpoints. A positive coefficient indicates the sensor signal
increases with increasing temperature. A negative coefficient indicates the sensor signal decreases
with increasing temperature.
Table 5-2. Recommended Curve Parameters
Type
Typical Lake
Shore Model
Units
Format
Limit
(K)
Coefficient
Recommended Sensor
Resolution
Silicon Diode
DT-470
V
V/K
475
Negative
0.00001 (V)
GaAlAs Diode
TG-120
V
V/K
325
Negative
0.00001 (V)
Platinum 100
PT-100
/K
800
Positive
0.001 (
)
Platinum 1000
PT-100
/K
800
Positive
0.01 (
)
Rhodium-Iron
RF-100
/K
325
Positive
0.001 (
)
Carbon-Glass
CGR-1-1000
log
/K
325
Negative
0.00001 (log
)
Cernox™
CX-1030
log
/K
325
Negative
0.00001 (log
)
Germanium
GR-200A-100
log
/K
325
Negative
0.00001 (log
)
Rox™
RX-102A
log
/K
40
Negative
0.00001 (log
)
Type K
9006-005
mV
mV/K
1500
Positive
0.0001 (mV)
Type E
9006-003
mV
mV/K
930
Positive
0.0001 (mV)
Type T
9006-007
mV
mV/K
673
Positive
0.0001 (mV)
Au-Fe 0.03%
No Longer Sold
mV
mV/K
500
Positive
0.0001 (mV)
Au-Fe 0.07%
9006-001
mV
mV/K
610
Positive
0.0001 (mV)
Содержание 325
Страница 4: ...Lake Shore Model 325 Temperature Controller User s Manual...
Страница 6: ......
Страница 22: ...Lake Shore Model 325 Temperature Controller User s Manual 1 10 Introduction This Page Intentionally Left Blank...
Страница 47: ...Lake Shore Model 325 Temperature Controller User s Manual 3 11 This Page Intentionally Left Blank...
Страница 48: ...Lake Shore Model 325 Temperature Controller User s Manual 3 12 This Page Intentionally Left Blank...
Страница 76: ...Lake Shore Model 325 Temperature Controller User s Manual 4 28 Operation This Page Intentionally Left Blank...
Страница 122: ...Lake Shore Model 325 Temperature Controller User s Manual 6 36 Remote Operation This Page Intentionally Left Blank...
Страница 128: ...Lake Shore Model 325 Temperature Controller User s Manual 7 6 Options and Accessories...