Lake Shore Model 325 Temperature Controller User’s Manual
6-26
Remote Operation
CRVHDR
Curve Header Command
Input:
CRVHDR <curve>, <name>, <SN>, <format>, <limit value>,
<coefficient>[term]
Format:
nn,aaaaaaaaaaaaaaa,aaaaaaaaaa,n,±nnn.nnn,n
<curve>
Specifies which curve to configure. Valid entries: 21
–
35.
<name>
Specifies curve name. Limited to 15 characters.
<SN>
Specifies the curve serial number. Limited to 10 characters.
<format>
Specifies the curve data format. Valid entries: 1 = mV/K, 2 = V/K, 3 =
/K,
4 = log
/K.
<limit value>
Specifies the curve temperature limit in kelvin.
<coefficient>
Specifies the curves temperature coefficient. Valid entries: 1 = negative,
2 = positive.
Remarks:
Configures the user curve header.
Example:
CRVHDR 21,DT-470,00011134,2,325.0,1[term]
– Configures User Curve 21 with a name of DT-
470, serial number of 00011134, data format of volts versus kelvin, upper temperature limit of 325 K.
The coefficient parameter does not actually set the temperature coefficient. It is only a placeholder so
that the CRVHDR command parameters match the CRVHDR? query parameters. The temperature
coefficient is determined by the first two points in the curve.
CRVHDR?
Curve Header Query
Input:
CRVHDR? <curve>[term]
Format:
nn
<curve>
Valid entries: 1
–
36.
Returned:
<name>, <SN>, <format>, <limit value>, <coefficient>[term]
Format:
aaaaaaaaaaaaaaa,aaaaaaaaaa,n,±nnn.nnn,n (Refer to command for description)
CRVPT
Curve Data Point Command
Input:
CRVPT <curve>, <index>, <units value>, <temp value>[term]
Format:
nn,nnn,±nnnnnnn,±nnnnnnn
<curve>
Specifies which curve to configure. Valid entries: 21
–
35.
<index>
Specifies the points index in the curve. Valid entries: 1
–
200.
<units value>
Specifies sensor units for this point to 6 digits.
<temp value>
Specifies the corresponding temperature in kelvin for this point to 6 digits.
Remarks:
Configures a user curve data point.
Example:
CRVPT 21,2,0.10191,470.000,N[term] –
Sets User Curve 21 second data point to 0.10191 sensor
units and 470.000 K.
CRVPT?
Curve Data Point Query
Input:
CRVPT? <curve>, <index>[term]
Format:
nn,nnn
<curve>
Specifies which curve to query: 1
–
35.
<index>
Specifies the points index in the curve: 1
–
200.
Returned:
<units value>, <temp value>[term]
Format:
±nnnnnnn,±nnnnnnn
(Refer to command for description)
Remarks:
Returns a standard or user curve data point.
Содержание 325
Страница 4: ...Lake Shore Model 325 Temperature Controller User s Manual...
Страница 6: ......
Страница 22: ...Lake Shore Model 325 Temperature Controller User s Manual 1 10 Introduction This Page Intentionally Left Blank...
Страница 47: ...Lake Shore Model 325 Temperature Controller User s Manual 3 11 This Page Intentionally Left Blank...
Страница 48: ...Lake Shore Model 325 Temperature Controller User s Manual 3 12 This Page Intentionally Left Blank...
Страница 76: ...Lake Shore Model 325 Temperature Controller User s Manual 4 28 Operation This Page Intentionally Left Blank...
Страница 122: ...Lake Shore Model 325 Temperature Controller User s Manual 6 36 Remote Operation This Page Intentionally Left Blank...
Страница 128: ...Lake Shore Model 325 Temperature Controller User s Manual 7 6 Options and Accessories...