80
c
HAPTER
4:
Operation
Model 372 AC Resistance Bridge and Temperature Controller
D
Locations:
depending on the number of displays parameter, there can be any-
where from two to eight display locations for displaying sensor readings. The
placement of a given display location on the front panel display depends on the
number of displays setting.
D
Number of custom display locations:
the number of displays parameter determines
how many sensor readings are displayed, as well as the character size of the dis-
played readings. If “2 (Large)” is selected, then two large character readings are
displayed, along with sensor names. If “4 (Large)” is selected, then four large
character readings are displayed, without sensor names. If “8 (Small)” is selected,
then eight small character readings are displayed, without sensor names.
Menu navigation:
Display Setup
Q
Number of Locations
(2 Large, 4 Large, 8 Small)
D
Input and units:
each available display location has an associated Input, and Units
setting. The input parameter determines which sensor will be used as the input
of the displayed data. The input can be the control input, measurement input
(and corresponding channels if a scanner is attached), or none. If none is selected,
then the display location will be blank. The units parameter determines which
units to display the reading in.
Menu navigation:
Display Setup
Q
Location
(1, 2, 3, 4, 5, 6, 7, 8)
Q
Input (None, Input A, 1-16)
Display Setup
Q
Location
(1, 2, 3, 4, 5, 6, 7, 8)
Q
Units (Kelvin, Ohms, Quadrature
[Measurement Input only],
Min, Max, Sensor Name)
Interface command:
DISPFLD
4.3.3 Control and Scan
Information Display
Quadrant
Independent of the display mode, the lower left quadrant of the display shows
relevant control loop information for either the sample heater or warm-up heater,
independent of display mode. In addition, it can be configured to display the active
measurement input scan channel.
FIGURE 4-7
Model 372 screen images showing
2 (top), 4 (middle) and 8 (bottom) display locations