96
c
HAPTER
4:
Operation
Model 372 AC Resistance Bridge and Temperature Controller
Default: 20 s
Interface command:
PID
4.6.1.5.4 Derivative (D)
The derivative parameter (also called rate) is the D part of the PID equation. It has a
setting range of 1 to 2500 s. A setting of zero turns D off. The default value for D is zero
(off).
To set D, first configure the output mode of the desired heater to Closed Loop PID
mode (section 4.6.1.4). Then press
Sample Heater
or
Warm-Up Heater
to display the
respective control loop information if it is not already displayed. Then press
PID/Man
Out
. This will cause a temporary Control Setup menu to be displayed allowing the D
value to be set.
Menu navigation:
Sample Heater
or
Warm-Up Heater
Q
PID/Man Out
Q
Derivative (D)
Q
(1 to 2,500)
Default: 0 s
Interface command:
PID
4.6.1.5.5 Manual Output
Manual Output is a manual setting of the control output. It can function in two differ-
ent ways depending on control mode. In open loop control mode, the Manual Output
is the only output to the load. You can directly set the control output from the front
panel or over the computer interface. In closed loop control mode, Manual Output is
added directly to the output of the PID control equation. In effect, the control equa-
tion operates about the Manual Output setting.
Manual output units follow the heater output display units of current or power. Refer
to section 4.6.1.3 to set the Heater Out display units. When heater display units are
configured as current, the manual output parameter is set in percent of full scale cur-
rent for the selected heater range for both the warm-up heater and the sample
heater. The Manual Output setting range is 0% to 100% with a resolution of 0.01%.
When the heater output display is configured for power, the warm-up heater manual
output is a percentage of full scale power. The Manual Output current setting range is
0% to 100% with a resolution of 0.01%. However, when configured for power, the
sample heater manual output parameter is set in watts, not percentage. An accurate
heater resistance must be entered under the sample heater output setup menu for
the power calculation to be accurate. The setting range is from 0 to full scale power
for the selected heater range with a setting resolution of six digits. The actual resolu-
tion is limited to the 0.01% resolution of the heater output current source.
Available full scale current and power are determined by the heater resistance, Max
Current setting, and Heater Range.
To set Manual Output, press
Sample Heater
or
Warm-Up Heater
to display the
respective control loop information if it is not already displayed, then press
PID/Man
Out
button. This will cause a temporary Control Setup menu to be displayed allowing
the I value to be set.
In addition, when an output is configured for Open Loop mode, the Manual Output
setting is available in the Output Setup menu.
Menu navigation:
Sample Heater
or
Warm-Up Heater
Q
PID/Man Out
Q
Manual Output
Q
(0% to 100%)