6.6.1 Interface Commands
165
|
www.lakeshore.com
INNAME
Sensor Input Name Command
Input
INNAME <input/channel>,<name>[term]
Format
n,s[15]
<input/channel>
Specifies which input or channel to configure: A (control input),
1 to 16 (measurement input)
<name>
Specifies the name to associate with the sensor input.
Example
INNAME 1,“Sample Space”[term]
— The string “Sample Space” will appear on the
front panel display, when possible, to identify the sensor information being displayed.
Remarks
Be sure to use quotes when sending strings, otherwise characters such as spaces and
other non alpha-numeric characters will be interpreted as a delimiter, and the full
string will not be accepted. It is not recommended to use commas or semi-colons in
sensor input names as these characters are used as delimiters for query responses.
INNAME?
Sensor Input Name Query
Input
INNAME? <input>[term]
Format
a
<input>
Specifies input to query: A to D (D1 to D5 for 3062 option).
Returned
<name>[term]
Format
s[15] (refer to command for description)
INSET
Input Channel Parameter Command
Input
INSET <input/channel>,<disabled/enabled>,<dwell>,<pause>,
<curve number>,<tempco>[term]
Format
nn,n,nnn,nnn,nn,n
<input/channel>
Specifies which input or channel to configure: A (control
input), 1 to 16 (measurement input), 0 = all channels (mea-
surement input).
<disabled/enabled>
Specifies whether the input/channel is disabled or enabled:
0 = disabled, 1 = enabled.
<dwell>
Specifies a value for the autoscanning dwell time:
1 to 200 s. Not applicable for <input/channel> = A (control
input).
<pause>
Specifies a value for the change pause time: 3 to 200 s.
<curve number>
Specifies which curve the channel uses: 0 = no curve, 1 to 59
= standard/user curves.
<tempco>
Sets the temperature coefficient that will be used for
temperature control if no curve is selected: 1 = negative,
2 = positive.
Example
INSET 5,1,10,3,0,2[term]
— Enables channel 5 with 10 s autoscan dwell, 3 s pause
time, and no curve with a positive temperature coefficient.
Remarks
At least one measurement input channel must be enabled. If all are configured to dis-
abled, channel 1 will change to enabled.
INSET?
Input Channel Parameter Query
Input
INSET? <input/channel>[term]
Format
nn
<input/channel>
Specifies which input or channel to query: A (control input),
1 to 16 (measurement input).
Returned
<off/on>,<dwell>,<pause>,<curve number>,<tempco>[term]
Format
n,nnn,nnn,nn,n (refer to command for description)