4.6.3 Power Up Enable
103
|
www.lakeshore.com
4.6.2.2 Still Output Mode
The Model 372 can provide the small amount of power required to drive the still
heater of a dilution refrigerator with the analog output. None of the analog output
hardware changes when it is configured for still operation. Section 3.8.2 describes
how to estimate the full scale current sourced to a resistive heater. When the analog
output is in still mode, the
Still
direct operation key gives direct access to the still out-
put parameter, which is used to set a percentage of the full scale current. The still out-
put parameter is the same as the manual output parameter in open loop mode, but
the setting can be made without sequencing through the analog output menu. Uni-
polar operation is recommended for this application.
Menu navigation:
Output Setup
Q
Analog/Still
Q
Output Mode
Q
(Still)
Interface command:
OUTMODE
Menu navigation:
Still
Q
Still Output
Q
(–100% to 100%)
Interface command:
STILL
4.6.3 Power Up Enable
All configuration parameters of the Model 372 can be retained through a power cycle.
Some systems require that the heater range or analog output is turned off when
power is restored. The power up enable feature allows you to choose whether or not
the heater range or analog output is turned off each time the instrument power is
cycled. Set the power up enable parameter to off to ensure that the heater range or
analog output is turned off on power up. Set it to on to return the heater range or ana-
log output to its previous setting when power is restored.
Menu navigation:
Output Setup
Q
Sample Heater, Warm Up Heater,
or
Analog/Still
Q
Power Up Enable
Q
(Off or On)
Default: Off
Interface command:
OUTMODE
4.7 Interface
The Model 372 has three computer interfaces: IEEE-488, USB, and Ethernet. Only one
of these interfaces can be active at one time. Use the Interface menu to configure
which interface is active, and to configure the parameters related to the selected
interface.
Menu navigation:
Interface
Q
Enabled
Q
(USB, Ethernet, IEEE-488)
Default: USB
4.7.1 USB
The USB interface is provided as a convenient way to connect to most modern com-
puters, as a USB interface is provided on nearly all new PCs as of the writing of this
manual. The Model 372 USB driver, which must be installed before using the inter-
face (section 6.3.3), creates a virtual serial com port, which can be used in the same
way as a traditional serial com port. Refer to Chapter 6 for details on computer inter-
face operation.
Menu navigation:
Interface
Q
Enabled
Q
USB