6.6.1 Interface Commands
159
|
www.lakeshore.com
CRVHDR?
Curve Header Query
Input
CRVHDR? <curve>[term]
Format
nn
<curve>
Valid entries: 1 to 59.
Returned
<name>,<SN>,<format>,<limit value>,<coefficient>[term]
Format
s[15],s[10],n,+nnn.nnn,n (refer to command for description)
CRVPT
Curve Data Point Command
Input
CRVPT <curve>,<index>,<units value>,<temp value>,
(optional)<curvature value>[term]
Format
nn,nnn,+nnn.nnnE±nn,+nnn.nnnE±nn,+nnn.nnnE±nn
<curve>
Specifies which curve to configure. Valid entries: 21 to 59.
<index>
Specifies the points index in the curve. Valid entries: 1 to 200.
<units value>
Specifies sensor units for this point to 6 digits.
<temp value>
Specifies the corresponding temperature in kelvin for this
point to 6 digits.
<curvature value>
Specifies the curvature value scale factor used to calculate cubic
spline coeffecients to 6 digits.
Remarks
Configures a user curve data point. When in 370 emulation mode, the user curves are
defined as 1 to 20 for backwards compatibility. The <curvature value> field is only
required if the curve point being loaded is used for a cubic spline interpolation curve,
and can be omitted when sending linear interpolation curves.
Example
CRVPT 21,2,0.10191,470.000,N [term]
—sets User Curve 21 second data point to
0.10191 sensor units and 470.000 K.
CRVPT?
Curve Data Point Query
Input
CRVPT? <curve>,<index>[term]
Format
nn,nnn
<curve>
Specifies which curve to query: 1 to 59.
<index>
Specifies the points index in the curve: 1 to 200.
Returned
<units value>,<temp value>,(optional)<curvature value>[term]
Format
±nnn.nnnE±nn,+nnn.nnnE±nn,+nnn.nnnE±nn (refer to command for description)
Remarks
Returns a standard or user curve data point. The curvature value is only returned
when the curve point being queried is part of a cubic spline curve.
DFLT
Factory Defaults Command
Input
DFLT 99[term]
Remarks
Sets all configuration values to factory defaults and resets the instrument. The “99” is
included to prevent accidentally setting the unit to defaults.
Содержание 372
Страница 12: ...Model 372 AC Resistance Bridge and Temperature Controller ...
Страница 162: ...150 cHAPTER 6 Computer Interface Operation Model 372 AC Resistance Bridge and Temperature Controller ...
Страница 214: ...202 cHAPTER 8 Service Model 372 AC Resistance Bridge and Temperature Controller ...