5.4.1 Monitor Units
109
|
www.lakeshore.com
For example, if Polarity is set to Bipolar, then setting the Monitor Out –10 V parameter
to 0 K and the Monitor Out +10 V parameter to 100 K will cause the analog output to
correspond to the input temperature as shown in FIGURE 5-3. In this case if the actual
reading was 50 K, then the output would be at 0 V (middle of the scale).
If we set the Polarity parameter to Unipolar, the output would be as shown in
FIGURE 5-4. In this case if the actual reading was 50 K, the analog output would be
+5 V (middle of the scale).
Menu navigation:
Output Setup
Q
Analog/Still
Q
Polarity
Q
(Unipolar or Bipolar)
Output Setup
Q
Analog/Still
Q
Monitor Out
-10 V
Q
(See note below)
Output Setup
Q
Analog/Still
Q
Monitor Out
0 V
Q
(See note below)
Output Setup
Q
Analog/Still
Q
Monitor Out
+10 V
Q
(See note below)
Monitor Out -10 V, 0 V, and +10 V settings depend on the Monitor Units selected, and are
limited to the acceptable values of the selected units.
Default:
Polarity
Q
Unipolar
Monitor Out -10 V
Q
0.0000 K
Monitor Out 0 V
Q
0.0000 K
Monitor Out +10 V
Q
1000 K
Interface command:
ANALOG
FIGURE 5-3
Analog output with polarity set to bipolar
FIGURE 5-4
Output with polarity parameter set to unipolar
Bipolar
0 K
50 K
100 K
–10 V
0 V
+10 V
Input
Output
0 V
+5 V
+10 V
0 K
50 K
100 K
Input
Output
Unipolar