8.7.2 Product Information
187
|
www.lakeshore.com
8.7.2 Product
Information
Product information for your instrument is also found in the Factory Reset menu. The
following information is provided:
D
Firmware version
D
Firmware date
D
Serial number
D
Scanner type
D
Ethernet version
D
Board revision
8.8 Error
Messages
The following are error messages that may be displayed by the Model 372 during
operation.
8.9 Calibration
Procedure
Instrument calibration can be obtained through Lake Shore Service. Refer to
section 8.14 for technical inquiries and contact information.
Message
Description
CS_OVL
Measurement input current source overload possibly caused by a broken current
source lead.
VCM_OVL
Measurement input common mode overload at the first amplifier stage of the voltage
input.
VDIF OVL
Overload at the differential amplifier stage of the measurement input voltage input.
VMIX OVL
Overload at the mixer (phase sensitive detector) stage of the measurement input volt-
age input.
R.OVER
Input is at or over full-scale sensor units.
R.UNDER
Input is at or under negative full-scale sensor units.
T.OVER
Input at or over the high end of the curve.
T.UNDER
Input at or under the low end of the curve.
DISABLED
Input is disabled. Refer to section 4.4.
NOCURV
Input has no curve.Refer to section 4.4.12.
Cannot Communicate with
Input DSP
The main microprocessor has lost communication with the sensor input DSP.
A temperature limit has
been exceeded
The temperature reading on a sensor input has exceeded the Temperature Limit set-
ting. A detailed message will follow, which includes a reference to which sensor inputs
temperature limit has been exceeded.
*** Keypad Locked ***
An attempt has been made to change a parameter while the keypad is locked.
Refer to section 4.8.
*** Heater Short Circuit
Detected ***
A short circuit condition has been observed on the warm-up heater output. A detailed
message will follow. The output will be turned off when this occurs.
*** Heater Open Circuit
Detected ***
An open circuit condition has been observed on the warm-up heater output. A detailed
message will follow. The output will be turned off when this occurs.
*** Invalid Calibration ***
The calibration memory is either corrupt, or is at the default, uncalibrated state. This
message appears when the Model 372 is first powered on. To clear the message, and
continue with instrument start-up, press the Escape and Enter keys simultaneously.
*** Firmware Update in
Progress ***
The Model 372 is in firmware update mode.
***Output not configured
for still***
The
Still Output
shortcut key has been pressed and the analog/still output is not con-
figured for still mode.
TABLE 8-2
Error messages