Advanced Operation
Model 372 AC Resistance Bridge and Temperature Controller
5.5.2 Relays
There are two relays on the Model 372, numbered 1 and 2. They are most commonly
thought of as alarm relays, but they may also be manually controlled. Two relays can
be used with one sensor input for independent high and low operation, or each can be
assigned to a different input.
When using relays with alarm operation, set up alarms first. The relays are rated for
30 VDC and 2 A. Their terminals are in the detachable terminal block on the
Model 372 rear panel.
In the Off mode, the relay is un-energized, leaving the normally open (NO) contacts
open and the normally closed (NC) contacts closed. In the On mode, the relay is ener-
gized, so the NO contacts will be closed and the NC contacts will be open. In the Alarm
mode the relay will activate based on the state of the configured Alarm Input sensor.
When the Alarm to Follow parameter is set to Low, the relay will energize if the con-
figured Alarm Input sensor goes into a low alarm state. If it is set to High, the relay will
energize if the configured Alarm Input sensor goes into a high alarm state. If the
Alarm to Follow parameter is set to Both, the relay will energize if the configured
Alarm Input sensor goes into either a low alarm or a high alarm state.
In the zone mode, each relay can be configured to an on or off state that is dependent
on the current zone. Because the zones are a function of an output, the Output to Fol-
low parameter lets the Model 372 determine which output zone settings the relays
will follow. The relay 1 and 2 zone states can be configured in the zone settings menu
for the respective control output.
Menu navigation:
Relays/Alarms
Q
Relay Setup
Q
(
Relay 1, Relay 2
)
Q
Mode
Q
(Off, On, Alarm, Zone (Sam-
ple), Zone (Warm up))
Relays/Alarms
Q
Relay Setup
Q
(
Relay 1, Relay 2
)
Q
Alarm Input
Q
Input (A, 1–16)
Relays/Alarms
Q
Relay Setup
Q
(
Relay 1, Relay 2
)
Q
Alarm to Follow
Q
(Low, High, Both)
Default:
Mode
Q
Off
Alarm Input
Q
Channel 1
Alarm to Follow
Q
Both
Interface command:
RELAY