6.6.1 Interface Commands
157
|
www.lakeshore.com
ANALOG?
Analog Output Parameter Query
Input
ANALOG? <analog channel>[term]
Format
n
<analog channel>
Specifies analog output channel: 1 or 2.
Returned
<polarity>,<mode>,<input/channel>,<source>,<high value>,<low value>,<manual
value>[term]
Format
n,n,nn,n, ±nnn.nnnE±nn, ±nnn.nnnE±nn,±nnnnnn
(refer to command for definition)
AOUT?
Analog Output Data Query
Input
AOUT? <output>[term]
Format
n
<output>
Specifies which unpowered analog output to query: 1 or 2.
Returned
<output percentage>[term]
Format
±nn.nnn
Remarks
Returns the output percentage of the unpowered analog output.
BAUD
RS-232 Baud Rate Command
Input
BAUD <bps>[term]
Format
n
<bps>
Specifies Baud rate: 0 = 300 Baud, 1 = 1,200 Baud, 2 = 9,600
Baud, 3 = 57,600 (default)
BAUD?
RS-232 Baud Rate Query
Input
BAUD?[term]
Returned
<bps>[term]
Format
n (refer to command for description)
BEEP
Audible Alarm Beeper Command
Input
BEEP <state>[term]
Format
n
<state>
0 = off, audible alarm beeper remains off
1 = on, audible alarm beeper is on when any alarm is active
Remarks
Enables or disables system beeper sound when an alarm condition is met
BEEP?
Audible Alarm Beeper Query
Input
BEEP?[term]
Returned
<state>[term]
Format
n (refer to command for description)
BRIGT
Display Brightness Command
Input
BRIGT <bright>[term]
Format
n
<bright>
Specifies display brightness: 0 = 25%, 1 = 50%, 2 = 75%,
3 = 100%.
Remarks
Sets the brightness for the front panel vacuum fluorescent display.