168
c
HAPTER
6:
Computer Interface Operation
Model 372 AC Resistance Bridge and Temperature Controller
INTYPE?
Input Setup Query
Input
INTYPE? <input>[term]
Format
nn
<input/channel>
Specifies input to query: A = control input;
1 to 16 = measurement input channel
Returned
<mode>,<excitation>,<autorange>,<range>,<cs shunt>,<units>[term]
Format
n,nn,n,nn,n,n (refer to command for description)
KRDG?
Kelvin Reading Query
Input
KRDG?<input>[term]
Format
n
<input>
Specifies which input to query: A = control input;
1 to 16 = measurement input Channel
Returned
<kelvin value>[term]
Format
±nnn.nnnE±nn
Remarks
Also see the RDGST? query.
LEDS
Front Panel LEDs Command
Input
LEDS <off/on>[term]
Format
n
<off/on>
0 = LEDs off, 1 = LEDs on
Remarks
If set to 0, front panel LEDs will not be functional. Function can be used when display
brightness is a problem.
Example
LED 0[term]
—turns all front panel LED functionality off.
LEDS?
Front Panel LEDs Query
Input
LEDS?[term]
Returned
<off/on> [term]
Format
n (refer to command for description)
LOCK
Keypad Lock Command
Input
LOCK <state>,<code>[term]
Format
n,nnn
<state>
0 = Unlocked, 1 = Locked
<code>
Specifies lock-out code. Valid entries are 000 to 999.
Remarks
Locks out all front panel entries except pressing the
Alarm
key to silence alarms, or
selecting to turn the heater off.
Example
LOCK 1,123[term]
—enables keypad lock and sets the code to 123.
LOCK?
Keypad Lock Query
Input
LOCK?[term]
Returned
<state>,<code>[term]
Format
n,nnn (refer to command for description)