51
Step 6:
Pressing the
〔
〕
key saves the adjusted value and the LCD shows “SAVE” symbol in the upper
part for 2 seconds.
Step 7:
Pressing the
〔
ZERO
〕
key exits the CAL mode and back to step 4 for next range, until other range
adjustment is finished.
11.2 Adjusting Frequency
Step 1:
Make sure the lead cables for measurement are not connected to the measuring instrument under
test.
Step 2:
Using the
〔
FREQ
〕
key, select frequency function.
Step 3:
Connect the lead cables to the output terminals of the standard source as shown in Figure 16.
Step 4:
The measured value and unit shall be shown in the upper part of the LCD.
Step 5:
Pressing the
〔
ZERO
〕
key enters the measurement CAL mode. The LCD shows the present adjusting
point “CAL:0” in the lower part and the reference frequency and unit needed for the point in the upper part.
Step 6:
Pressing the
〔
〕
key saves the adjusted value and the LCD shows “SAVE” symbol in the upper
part for 2 seconds.
Step 7:
Pressing the
〔
ZERO
〕
key exits the CAL mode. The adjustment is finished.