19
7 Source
From the calibrator, you can source a DC voltage, DC current, frequency, pulse signal.
Warning
To avoid electrical shock, do not apply more than the rated voltage, as marked on the calibrator, between
terminals or between any terminal and earth ground. Always use the calibrator in locations with a voltage to
ground below 30 Vpk.
Caution
z
Do not apply any voltage to the output terminals for ranges other than 4-20mA simulating transmitter
output Otherwise, the internal circuitry may be damaged.
7.1 Connecting Cables to Terminals
For DC voltage, frequency, pulse (Figure 5)
Step 1:
Connect the black lead cable for source to the
COM output terminal and the red lead cable to the “VHz”
output terminal.
Step 2:
Connect the other ends of the cables to the input
of equipment under test while making sure the polarities
are correct. Figure 5 Sourcing DC voltage, frequency, pulse