12 Adjusting Source Functions
Environmental Requirements
Ambient temperature: 23
±
5
℃
Relative humidity: 35% to 75% RH
Warm-up:
z
Before using, warm up the calibrator for the period of time specified.
z
Set the meter into the standard environment for 24 hours, and then turn on the power. Change the set into
non-automatic power-off state and warm it up for one hour.
Power Supply: new alkaline size (AAA) battery type 7 is the best choice for adjustment.
Source Adjustment Operation:
Table 5 Adjustment Points of Source Functions
Range
Adjustment Point *
Remarks
0 A F
DCV_1000mV 0
0
1000mV
DCV_10V 0
-
10V
DCmA_20mA 0
-
20mA
* Adjusting the displayed value same with the reading of the digit meter when the present calibrator is
stabilized.
z
You can calibrate a desired function and range separately.