48
Tips:
Any change of setting to the above-mentioned function, press the
〔
ON
〕
key to save the value. Any press of
the
〔
ON
〕
key saves the nearest setting value.
11 Adjusting Measurement Functions
Environmental Requirements
Ambient temperature: 23
±
5
℃
Relative humidity: 35% to 75% RH
Warm-up:
z
Before using, warm up the calibrator for the period of time specified.
z
Put the meter into the standard environment for 24 hours, and then turn on the power. Change the set into
non-automatic power-off state and warm it up for one hour.
Caution:
Power Supply: new alkaline size (AAA) battery type 7 is the best choice for adjustment.
Содержание KM-CAL-907