41
Tips:
z
For absolute pressure module, the calibrator saves the zero-off value and reuses the value automatically.
Therefore, user does not need to zero off the calibration value for each use.
z
Pressing
the
〔
UNIT
〕
key convert pressure unit (kg/cm
2
, cmH
2
O, mmHg, PSI, bar, mbar, MPa, Kpa).
9.2 Calibrate Pressure-Voltage Transmitter
Calibrator can measuring voltage signal from transmitter under
pressure mode.
Step 1:
Connect the pressure module and calibrator as shown in
Figure 13.
Step 2:
Pressing the
〔
ON
〕
key, the calibrator connects and
senses the type of the pressure module and sets the range
automatically. If it fails to connect, the LCD shows “NO.OP” in
the upper part.
Step 3:
Press the
〔
V
〕
key to select DC voltage measurement
function.
Figure 13 calibrate pressure-voltage transmitter