40
Step 1:
Connect the pressure module and calibrator as shown in Figure 12. The screw of the pressure module
pipe is compatible to the 1/4 inch NPT connector. If you have other requirement, contact the vendor.
Step 2:
Using the
〔
MODE
〕
key, select the pressure measurement function. The LCD shows “PRESSURE” at
right upper part.
Step 3:
Pressing the
〔
ON
〕
key, the calibrator connects and senses the type of the pressure module and sets
the range automatically. If it fails to connect, the LCD
shows “NO.OP” in the upper part.
Step 4:
Zero off the reading following the pressure
module manual. When the reading overtops the 5
percentage of maxim value of the range, the LCD shows
“ERR” in the lower part. Pressing the
〔
ZERO
〕
key
initializes the calibrator to 0,and “
△
”symbol shows on the
left upper part of the LCD.
Figure 12 Measuring pressure